Beam-size measurement with optical diffraction radiation at KEK accelerator test facility

Karataev, Pavel, Araki, S , Hamatsu, R , Hayano, H , Muto, T , Naumenko, G , Potylitsyn, A , Terunuma, N and Urakawa, J

(2004)

Karataev, Pavel, Araki, S , Hamatsu, R , Hayano, H , Muto, T , Naumenko, G , Potylitsyn, A , Terunuma, N and Urakawa, J (2004) Beam-size measurement with optical diffraction radiation at KEK accelerator test facility. Physical Review Letters, 93 (24).

Our Full Text Deposits

Full text access: Open

Full text file - 354.72 KB

Abstract

An experiment on the investigation of optical diffraction radiation (ODR) from a slit target as a possible tool for noninvasive electron beam-size diagnostics has been performed at the KEK accelerator test facility. The experimental setup has been installed at the diagnostics section of the extraction line. We have performed the first incoherent ODR observation from a slit target. The measured angular distributions are in reasonable agreement with the theoretical expectation. The beam-size effect onto the ODR angular pattern has been observed. Moreover, the sensitivity to the beam size as small as 14 mum has been achieved.

Information about this Version

This is a Submitted version
This version's date is: 10/12/2004
This item is not peer reviewed

Link to this Version

https://repository.royalholloway.ac.uk/items/e178c091-50fc-2112-34e4-3769b3e56b7e/7/

Item TypeJournal Article
TitleBeam-size measurement with optical diffraction radiation at KEK accelerator test facility
AuthorsKarataev, Pavel
Araki, S
Hamatsu, R
Hayano, H
Muto, T
Naumenko, G
Potylitsyn, A
Terunuma, N
Urakawa, J
Uncontrolled KeywordsDIAGNOSTICS
DepartmentsFaculty of Science\Physics
Research Groups and Centres\Physics\Low Temperature Physics

Identifiers

doihttp://dx.doi.org/10.1103/PhysRevLett.93.244802

Deposited by Research Information System (atira) on 18-Nov-2014 in Royal Holloway Research Online.Last modified on 18-Nov-2014


Details